How to Initiate an Exetainer Sequence on NACHO

INTRODUCTION

This document describes how to initiate a run on the Delta V mass spectrometer (“NACHO”) in OSB 446 to analyze d18O-O2, O2/Ar, 15N-N2, and d13C-Dissolved Inorganic Carbon (DIC) in Exetainer headspace.  The protocol assumes samples and water standards have already been prepped for analysis following the Prepping Exetainers For Analysis protocol.  Also included is a basic description of raw data decomposition and standardization using a developed R script.

Note that the IsoDat software to interface with the mass spec has three versions, each with separate functions.  IsoDat Workspace is for manipulation files while the instrument is running. Acquisition is used to collect data.  IsoDat Instrument Control is used for changing settings and tuning. You should never need to use IsoDat Instrument Control.

SAFETY

You will be using needles and a slightly compressed gas.  If you are unfamiliar about how to work with compressed gases, you must first take the Online Compressed Gas Safety Course.  Samples may contain relatively dilute concentrations of zinc chloride (ZnCl2) and have been acidified to pH 2.  You should not come in contact with these chemicals under normal operation.  Please familiarize yourself with the MSDSs and SOPs for these chemicals nonetheless. Please also dispose of sharps responsibly.

MATERIALS

  • 8-12 Exetainers with new caps and 2-4 glass bead (for air standards)
  • ~1.7% CO2standard (balance He)
  • 5 mL syringe with 30 G needle
  • Lab timer
  • Vacuum oil and dedicated pipette
  • Small Erlenmeyer flask or beaker
  • Microspatula (not essential, but helpful)
  • Kim wipes

PREPARING AIR STANDARDS (OSB 435)

  1. Prepare air standards by placing a new cap + septa on 10-12 vials that each have 2-4 glass beads. Place in a wire rack.
  2. Take the vials to the CO2tank by Johnny’s office. Open the tank and then the small black knob on the regulator. Set the regulator to approximately 2 psi pressure.
  3. Insert the long, vent needle all the way to the bottom of the Exetainer. The other end of the tubing should be in a small container (Erlenmeyer flask or beaker) of water.
  4. Place the Exetainer upside down on the CO2needle with the needle outlet near the septa. Check for bubbling at a rate of 2-3 bubbles per second.
  5. After three minutes (use lab timer), remove the Exetainer from the CO2needle and wait until the vent needle stops bubbling, or approximately 10 seconds.Remove vent needle and return Exetainer to rack.
  6. Repeat steps 3-5 for each of the Exetainers.
  7. Take Exetainers to the fourth-floor outdoor balcony to fill with air over a range of volumes that reflects the expected range of O2concentration in the samples. A typically range for low oxygen systems (warm, heterotrophic) would be 0.2 – 0.6 cc of air. For samples from more oxic environments a range of 0.4-0.8 cc would be appropriate. Regardless of expected O2concentration, include standards at 0.2 and 0.8 cc air in every run to identify mass effects on 18O-O2and O2/Ar.  All standards should be made in duplicate.

CONFIGURING NACHO

  1. At the white pressure controller near the autosampler, check to see if the helium pressure is set to 23 psi. If it is not at 23psi when you switch to the Exetainer method, you will flush NACHO with air, which will take days to purge.
  2. In IsoDat Acquisition, select the pulldown menu at the bottom left that should read “GC-C Interface” and change to ‘Exetainers’ as the new method. The software will take a minute to reconfigure. (Figure 1, left side)
  3. It possible that a little bit of air will enter the system. Confirm any air has been purged by switching to the ‘N2’ gas configuration using the pulldown menu just to the right of one to set the system configuration. Under the MS window, mass 30 should be <100 mV (typically around 60 mV). You may also check background levels of other gases (O2/Ar, N2, and CO2) by toggling between those mass jumps. You may toggle between them using the pop-up menu in the lower left-hand corner of the application, next to the method indicator. (Figure 1, right side)
  4. Expand the ConFlo IV Diagnosis window on the left side of IsoDat Acquisition. Confirm that configure matches the picture below. (Figure 2)
Details about how to switch NACHO "method"
Figure 1.
Figure 2.

ORGANIZING YOUR RUN

  1. A typical run starts with three air standards, with the first discarded as a conditioner. The remaining standards are spread throughout the run, typically placed at the beginning of each row. Randomize the order with respect to amount of air.
  2. Arrange the air standards (n=9), water standards (n=2), and exetainer samples (n=31) on the sampling block. The sampling block positions are labeled with even numbers only. Starting at position #2:
    1. 3 air standards (first air standard is conditioner)
    2. 1 water standard
    3. Finish the first row with exetainer samples
    4. Place an air standard at the start of each subsequent row, followed by samples
    5. Second to last sample position should be the remaining water standard
    6. Final sample position is the final remaining air standard
      Figure 3. A bird’s-eye view of the sample block loaded with air and water standards only. The remaining empty positions may be filled with samples.
  3. Note: Be aware that the numbering on the sample block does skip some even numbers.
  4. Take the block to the computer station to enter the standard and sample information into a Sequence File.
    1. In IsoDat Workspace, open Exetainers_template.seg and “Save as…” a unique identifier for your run/sequence. This file has spaces for a full run.
    2. In the column labeled “Identifier 1”, enter a unique name for each sample and/or standard replicated 4 times. For example, the first four lines should likely read “airSTD_0” or similar. Every Exetainer is measured 4 times (thus the 4 rows). Each Exetainer must have a unique ID, even air standard replicates (Ex: airSTD_0.2_R1, airSTD_0.2_R2).
    3. In the column labeled “Comment”, identify each Exetainer as either a “Conditioner”, “Standard”, or “Sample”, again replicating 4 times. Use only these three options. Water standards may be labeled as “Standard.”
    4. In the “Identifier 2” column, add the amount of air added to each standard.This column is only used for standards and must be numeric (i.e., no ‘cc’ at the end).
    5. Go back through the sequence file to make sure no rows are skipped and that the sample information lines up correctly with the AS field that identifies the position in the block. Also check that the ‘Amount’ field contains the sequence 1,2,2,2 for each Exetainer.A value of 1 tells the autosampler to move to that position.
    6. Save and close the sequence file.
  5. For each sample and water standard, thoroughly remove all the Apiezon grease on the septa using Kim wipes or paper towels. The grease will clog the needle. Pro tip: a microspatula is useful to efficiently remove all Apiezon grease.
  6. Add enough new vacuum oil to the septa of every Exetainer to fill the indented part of the cap using the dedicated pipette. The vacuum oil seals the septa and lubricates the needle. Don’t be stingy.
  7. Place the block in the autosampler tray making sure to engage the alignment tabs.
  8. Turn on the autosampler (black switch on rear) while holding on the Exetainer where the needle is resting (important, or the exetainer will go flying across the room).The autosampler should lift the needle up then undergo a quick self-check. Place the spare Exetainer out of the way of the autosampler.

STARTING THE RUN

  1. Reopen working sequence file in IsoDat Acquisition.
  2. If the mass spec run will include anything less than the full autosampler tray (i.e., running a subset of samples or a small sequence), highlight all the rows corresponding to the samples in the block.
  3. Make sure that all of your desired sample rows are highlighted before you start the machine. NACHO will only run what has been highlighted in IsoDat.
  4. From the drop-down menus at the top of the screen select AcquisitionàStart.Do not hit the “start button” above the spreadsheet unless you have a full block and all rows in the sequence file correspond to samples.
  5. An acquisition window will appear. The application will ask you to name the output .xlsx. Delete the word “acquisition” in the file name for brevity. Specify the export file type as an Excel file and rename the export file, ending with the date in the format yyyymmdd.
  6. Click OK to begin the run.
  7. Check that the autosampler needle has fully penetrated the septum of the first vial.The hexagonal nut above the needle should be positioned below a white plate. If it is not, you can push the needle down with your finger.  You may have to apply more vacuum oil to prevent this problem from affecting further samples.
  8. It is advisable to watch the first couple of sample injections to make sure everything is working properly. Specifically, you are looking for the presence of 7 peaks of roughly equal size and the absence of a small pressure peak immediately after the first two reference gas peaks. See example chromatograms.

CHECKING YOUR RUN DATA

In the middle of your run or especially at the end, there are a couple of things you can check to make sure that your sequence has good data. Generally, you may disregard the data from each first injection of a standard or sample.

  1. Look at the d34/32O2of the air standards. They should range from -2.5 to -1.5. The first injection is often up around -3, but you can disregard that.
  2. Check that the d13C is around -26.6 for injections 2-4. This value is relative to the working standards, so don’t expect them to match known values.
  3. Check that the area for the d32-O2curves at high and low standard concentrations bracket the d32-O2areas for your unknown samples. Again, look at the 2ndor 3rdinjections across vials. If they do not, you will need to run additional air standards with a different air range in the next batch of samples. For example, if your samples have more O2than your highest standards, add standards with more air in them.

ENDING A RUN & CLEAN UP

  1. Remove standards and samples from the block. Place in a cardboard Exetainer box or rack and return to the HEEL for cleaning and reuse.
  2. The system is ready to start another run if desired (return to beginning of this protocol).Follow the remaining steps if the system will remain idle for more than a couple of hours.
  3. Place an empty Exetainer (with cap) under the needle at its home position. Turn off the autosampler. The needle will drop down on to the Exetainer but not puncture the septa. Align the septa under the needle and push the autosampler needle through.
  4. Lower the He carrier gas pressure (white box) from 23 psi to 5 psi. Wait to make sure the pressure is stable at 5 psi and does not go to zero.
  5. In IsoDat Acquisition, switch from the “Exetainer’ configuration (bottom left pull down menu) to the “GC-C Interface” configuration.

RETREIVING the DATA

Option 1

  1. On the Desktop, open the EA_Results folder and select the date of your run. These are saved using the nomenclature yymmdd. Check the chromatograms (.dmf files) to ensure the presence and adequate separation of all peaks. Check within all folders with the dates that apply to your mass spec run.  If the IsoDat Acquisition was restarted due to a communication error between the IsoDat Acquisition and the Delta V mass spec, this will lead to multiple Excel summary files pertaining to your run. Copy each one to your hard drive.
  2. Save the data to a disc. Do not transfer files using a USB; this may introduce viruses to this PC.

Option 2 (preferred)

  1. Data are dynamically backed up to the HEEL NAS system and can be accessed there.First, network to the following location:
    1. //acoustics.washington.edu/home/CloudStation/Backup/nacho_OSB446_DellPC/G9J8G32/C/Thermo/Isodat NT/Global/User/Conflo II Interface/Results
    2. The user is HEEL, password: TreyReil1995.
  2. Check within all folders with the dates that apply to your mass spec run. If the IsoDat Acquisition was restarted due to a communication error between the IsoDat Acquisition and the Delta V mass spec, this will lead to multiple Excel summary files pertaining to your run.Copy each one to your hard drive.

HELPFUL TIPS

HOW TO UNCLOG THE AUTOSAMPLER NEEDLE

  • Ensure that the autosampler needle is not clogged before beginning each run. The autosampler needle will need to be unclogged if a small pressure peak appears after the second O2/Ar reference peak.
  • To unclog, first check for the He stream by submerging the autosampler needle in water.
  • Remove from water, wipe with Kim wipe, and gently clear the first 1-2 mm of the inside of the bottom opening of any septa fragments with a small gauge needle or piece of wire.
  • Increase the pressure of the He stream from its operational 14 psi to the maximum psi using the black knob at the back of the autosampler to purge. Cover the bottom opening on the autosampler needle with a finger to purge the top (intake) opening, wipe with a Kim wipe, and repeat.
  • Re-adjust the pressure to of the He stream to 14 psi.

EXAMPLE CHROMATOGRAMS

Your chromatogram should have 7 peaks, shown below, in the following order:

  • O2/Ar working standard reference peak
  • O2/Ar working standard reference peak
  • O2/Ar in your sample

**Mass jump**

  • N2in your sample
  • N2 working standard reference peak

**Mass jump**

  • CO2 in your sample
  • CO2 working standard reference peak
example Exetainer chromatograph
Figure 4.
  • If you notice a right skew in your peaks, you may need to condition or bake the column to remove any contamination. Skewed peaks will prevent complete peak integration before the machine moves on to the next mass jump, rendering your data unusable.
  • Working standards for O2/Ar (a custom, pre-mixed tank), N2and CO2come from standard gas tanks located in the closet just beyond the computer. These tanks have a long life on this machine since we draw microliters at a time.
  • The order and number of working standard injections is by the elution times of the sample gases. There is no oven on this column, so we cannot control when the sample gases elute from the column; we can only operate within their time frame.
  • We use two O2/Ar standards because 1) there is ample time before the O2/Ar sample elution and 2) it primes the machine a bit.
  • Sample data is calculated based on working standard data, which are verified by the air standards prepped at the beginning of this protocol.

WASTE

After your samples have been run, they need to be disposed of as chemical waste. Exetainer samples may contain relatively dilute concentrations of zinc chloride (ZnCl2) and have been acidified to pH 2 with 50% w/v H3PO4. Exetainer caps may be thrown away; they are not reused. Collect all liquid waste in a hazardous waste container (label suggestion below). Exetainers should be cleaned according to the “Exetainer Cleaning” protocol.

Chemical Composition %
Water 99
H3PO4 <1
ZnCl2 <1

Download a printable version of this protocol.

Authors: Ben Miller & Gordon Holtgrieve

Revision Date: 24 October 2018

Revision Author: Julia Hart